Automatic Systems Laboratories

page location: home > precision thermometers > f252

Precision Thermometers

F252 Precision Thermometer

Features

  Bullet

Accuracy: ±0.01°C; up to ±0.02°C with probe

  Bullet

4 wire AC bridge for stability without recalibrating

  Bullet

User selectable resolution of up to 0.001°C or 0.0001 ohms over full range

  Bullet

Range: -200°C to +962°C depending on probe

  Bullet

2 channels as standard with optional 4 or 6 channels expandable up to an additional 64 channels with Switchboxes

     

Typical Applications

  Bullet

Site Standard

  Bullet

Environmental Chambers

  Bullet

Research

  Bullet

Viscometry

  Bullet

Multi-channel logging

 

F252 BrochurePDF Document (PDF, approx 168KB)

 

For more information about this product contact ASL

 

F252 Digital Thermometer
Dotted line

ASL's F252 provides high accuracy, dual channel temperature measurement with Platinum Resistance Thermometers - PRTs - and uses AC bridge technology as developed by ASL for Standards Laboratories throughout the world. Model F252 requires no regular recalibration, the AC circuit techniques eliminating thermal EMFs and drift, while providing excellent long term stability. Front panel data entry allows you to enter probe calibration data into the F252's memory, so that thermometer resistance can be accurately converted to temperature in °C, °F or K. This can be password protected to prohibit unauthorised or accidental data entry. The combination of proven technology and sophisticated microprocessor control gives you a reliable, easy to use precision instrument.

Using F252 with SB252 Switchbox

Dotted line

The 8 or 16 channel SB252 switchbox complements the F252 in appearance and performance. Adding no extra system errors you can use one to four switchboxes connected to an expansion port on the rear of the F252, with either manual switching from the keypad or under interface control. With the addition of four 16 channels SB252's this can provide 64 additional channels to the F252. This turns the F252/SB252 into a powerful logging system when used with ASL's U-LOG data logging software.

F252 Precision Thermometer Features

Dotted line

See the table below for a run-down of this product's features and capabilities.

Range

-200°C to +962°C, depending on PRT used

Accuracy

F252 only:

±0.005°C (±5mK) at 0°C, otherwise <±0.01°C (±10mK)

F252 + T100-250:

±25mK (-50°C to +250°C)

F252 + T100-450:

-70°C (±45mK), -40°C to +250°C (±25mK),+250°C to +450°C (±100mK)

F252 + T100-650:

-189°C to 420°C (±20mK), +420°C to +550°C (±35mK)

Resolution

0.001°C /°F /K (user selectable)

0.0001ohms (user selectable)

Repeatability

±2 least significant digits (±0.002°C), included in accuracy

Stability

Long term: typically <0.002ohms per year (<0.005°C for Pt100)

Temperature co-efficient

<±0.0005°C/°C ambient change. Operating ambient: 15°C to 35°C (15°C to 25°C for best accuracy). 10-90% Rh non-condensing

Data entry format

ITS90 coefficients, Callender van Dusan coefficients or EN60751 for un-calibrated probes

Probe Current

1 mA constant current source

Probe Types

1 mA constant current source

Thermometer input connectors

Front panel:

2 x 5 pin industrial DIN socket

Rear panel:

Optional 2 or 4 x 5 pin industrial DIN socket

Communications

USB (standard)

Optional RS232C or IEEE-488 or LAN Ethernet

     

For more information about this product contact ASL

 